DEXA scan for body fat: what it measures

The words DEXA scan above a scanning table with an arm over it, beside a report listing Fat, Lean and Bone

A dual-energy X-ray absorptiometry (DEXA or DXA) scan uses two X-ray energies to split the body into fat, lean soft tissue and bone mineral, for the whole body and for the arms, legs, trunk and head. Researchers use it as the yardstick for body fat equations such as Relative Fat Mass, but its results shift between machine makers and models, with body thickness and with hydration. One whole-body scan gives an adult an effective dose of a few microsieverts, which a 2017 review compares to about one day of natural background radiation.

A DEXA scan, written DXA in research papers, is short for dual-energy X-ray absorptiometry. For body fat, it gives you three numbers for your whole body and for each region: how much is fat, how much is lean soft tissue, and how much is bone mineral.

It is also the yardstick that Relative Fat Mass, one of the equations on this site, was built against.

What a DXA scan measures

A DXA machine passes X-rays at two energies through your body while you lie on a table. A 2017 review in the journal Bone by John Shepherd and colleagues at the University of California, San Francisco, explains the principle: the two X-ray attenuations (how much of each beam the body absorbs) can be used to calculate the mass of two different materials, using simple algebra and the known physical properties of those materials.

The scanner then divides the body with set cut lines into arms, legs, trunk and head. The same review describes an early system, the Lunar DPX, that already reported fat, lean soft tissue and bone mass for each of those regions.

One detail matters when you compare a scan with a calculator. DXA reports lean soft tissue and bone mineral as two separate numbers. The lean body mass calculator on this site counts everything that is not fat, bone included. If you check a scan against that page, add the scan's bone figure to its lean figure first.

Why researchers use DXA to check body fat equations

The same 2017 review gives three reasons DXA is used for bone and body composition:

  • few assumptions: the two-beam physics does the separation
  • regional results: the cut lines give arm, leg and trunk values
  • stability: using test objects called phantoms, a single DXA system can be checked to stay stable to better than 0.5% over decades

DXA also sits in large data sets, and that is where Relative Fat Mass comes from. Orison Woolcott and Richard Bergman, at Cedars-Sinai Medical Center in Los Angeles, used the US National Health and Nutrition Examination Survey (NHANES) 1999 to 2004 (12,581 adults) to build the equation and NHANES 2005 to 2006 (3,456 adults) to test it. Their reference for "true" body fat was the DXA scan. The result, published in Scientific Reports in 2018, is 64 minus 20 times height divided by waist, plus 12 for women. In the test group it estimated DXA body fat more closely than body mass index (BMI) did, for both women and men.

Relative Fat Mass is one of the methods behind the body fat calculator. The other equations on that page were fitted against other lab methods, which is part of why body fat calculators disagree.

DXA is not the last word

Woolcott and Bergman name the problem in their own limitations. Compared with the four-compartment method, they write, DXA underestimates fat percentage at lower body fat and in men, and overestimates it at higher body fat and in women. So Relative Fat Mass could be slightly more or less accurate than they report, depending on body fat and sex.

A four-compartment model splits the body into fat, water, mineral and the rest, and measures several of these separately. In a 2017 study in the Journal of Clinical Densitometry, Watson, Venables and Murgatroyd built theirs from body weight, total body water (by deuterium dilution, a heavy-water test), body volume (by air displacement, the BOD POD method) and bone mass (by DXA). Watson 2017 and LaForgia 2009, both below, judge DXA against a four-compartment model.

Where DXA numbers drift

Machine makers and models

The International Society for Clinical Densitometry (ISCD), in its 2023 adult official positions, says comparing results across manufacturers needs an in vivo cross-calibration study (scanning the same people on both machines), and that no phantom has been found that removes those differences.

Four studies put a size on the gap:

StudyWhat was comparedWhat it found
Shepherd et al., 2012, Journal of Bone and Mineral Research199 people on both a GE Healthcare Lunar and a Hologic systemPercent fat differed by 1.4% on average (absolute difference)
Watson et al., 2017, Journal of Clinical Densitometry69 adults on two GE Lunar models, Prodigy and iDXA, plus a four-compartment modelThe Prodigy read more fat and less lean mass than the iDXA; the iDXA was closer to the four-compartment model, but a gap remained
Schoeller et al., 2005, American Journal of Clinical Nutrition1,195 adults on a Hologic QDR 4500A against criterion methodsIt overestimated fat-free mass by about 5% on average
Krueger et al., 2016, Journal of Clinical Densitometry30 volunteers on an old and a new Lunar iDXA, the same modelPercent fat differed by 1.1% on average, so the new machine was recalibrated

Software versions were a smaller issue in the Watson study: it found no significant difference in fat mass between the versions it compared.

So two scans compare best when they come from the same machine, or from machines that have been cross-calibrated.

Body size

Shepherd and colleagues describe two problems with large bodies: the person can be longer or wider than the scan area, and heavier than the table's rated load. The ISCD lists weight over the table limit as a limitation of clinical DXA for body composition, and says offset scanning should be used, with a procedure validated for that scanner model, for people too wide for the scan area. In offset scanning, one side of the body is scanned and mirrored to estimate the other (Shepherd 2017).

Body thickness also affects accuracy. In a 2009 study in Obesity, LaForgia and colleagues scanned 14 adults with a BMI of 28.7 to 39.9 and compared DXA with a four-compartment model. The group averages matched (41.1% against 41.5% body fat), but individual results were off by up to about 4 percentage points either way, and the error grew with tissue thickness such as chest depth.

Hydration

Changes in body fluid can shift a DXA result. Pietrobelli and colleagues, in a 1998 modelling study in the American Journal of Physiology, showed that extra body fluid produces small but systematic errors in DXA percent fat. It was a simulation, not a study of real people. In the LaForgia study, the differences between DXA and the four-compartment model were not linked to the hydration of fat-free mass.

The ISCD asks for consistent preparation between scans, naming fasting state, clothing, time of day, physical activity and an empty bladder.

The radiation dose

DXA uses ionising radiation, but the dose is small.

  • The 2017 Shepherd review gives the effective dose from a single whole-body DXA as under 10 microsieverts, similar to the natural background radiation received over one day at sea level.
  • A 2010 review in European Radiology by Damilakis and colleagues reports adult effective doses for whole-body DXA of 4.2 microsieverts on a Hologic Discovery A and 8.4 microsieverts on a Hologic Discovery W. The same review gives the worldwide average natural background dose as 2.4 millisieverts a year, which works out to about 6.6 microsieverts a day.
  • Damilakis and colleagues add that dose varies between models and makers.

The ISCD lists pregnancy as a contraindication to DXA body composition.

Common questions

Is a DEXA scan more accurate than a body fat calculator?

They answer different questions. A DXA scan measures fat, lean tissue and bone with X-rays, and Woolcott and Bergman used it as the reference when they built Relative Fat Mass from height and waist. DXA itself still differs from four-compartment research models, by up to about 4 percentage points for an individual in a 2009 study of adults with a BMI of 28.7 to 39.9.

Why did two DEXA scans give me different body fat results?

Results shift between manufacturers and even between two units of the same model. A 2012 study found an average absolute difference of 1.4% in percent fat between GE Lunar and Hologic scanners, and a 2016 study found a 1.1% gap between an old and a new scanner of the same model. Body fluid can shift the result too, and the ISCD asks for consistent preparation between scans: fasting state, clothing, time of day, physical activity and an empty bladder.

How much radiation is in a DEXA body fat scan?

A 2010 review in European Radiology reports 4.2 to 8.4 microsieverts for an adult whole-body scan on two Hologic models. A 2017 review in Bone puts a single whole-body scan under 10 microsieverts, similar to one day of natural background radiation at sea level.

Does a DEXA scan measure muscle?

It measures lean soft tissue: all soft tissue that is not fat. Muscle is part of that number, but the scan does not report muscle on its own. It reports bone mineral as its own number.

Sources

Every figure above comes from one of these. Check them.

Last reviewed 10 November 2026. Not medical advice.

Every method, with its source attached

The app runs body mass index (BMI), waist ratios and three body fat methods side by side, and tells you when they disagree about you.

Free, no account, and it works offline. 100,000+ downloads on Google Play.

This is not medical advice. Body Numbers is a calculator. It does not diagnose, treat or monitor any condition, and no number it produces should replace a conversation with a doctor or a dietitian. If a result worries you, or if you are pregnant, unwell, or under eighteen, speak to a health professional.